|
SEP-10S PLUS
Flow rate range:
0.1 - 99.9 ml/h in 0.1 ml/h steps
100 - 1500 ml/h in 1 ml/h steps
Mechanical accuracy:
±1%
Volumetric accuracy with approved
syringes: ±2% or ±1 ml/h
(the greater of these values)
Compatible syringes, ml:
10, 20, 30, 50/60, 100 ml more than 30 brands (BD, Monojet, Terumo,
BBraun, Fresenius, JMS and others)
Occlusion pressure:
40 kPa, 80 kPa, 120 kPa
Purge/Bolus rate 0;
10-1500 ml/h in 1 ml/h steps
Bolus volume:
0.1 - 99.9 ml in 0.1 ml steps
Keep Vein Open (KVO) rate:
0.1 - 10 ml/h in 0.1 ml/h steps
KVO volume:
1% of the syringe volume
Standby time:
1 - 999 min in 1 min steps
Volume infused:
0.1 - 99999 ml
Volume to be infused (VTBI):
0.1 - 99.9 ml in 0.1 ml steps
100 - 999 ml in 1 ml steps or unlimited volume
Alarms:
No Mains
Low Battery
Very Low Battery
Occlusion
x min Pre-alarm
KOR (KVO) xx ml/h
End of Infusion
Syringe Empty
Illegal syringe
Plunger Not Fitted
Clamp Opened
Internal Error
Selectable options for customized
operations: Enabled via set-up menu:
- VTBI
- Volume over time
- Infusion rate limit: xxxx ml/h
- Automatic/manual bolus
- Bolus rate programming
- Bolus rate limit: xxxx ml/h
- Total volume display
- Autosave
- Drug name display
- Titration
- Stanby
- Keypad lock
- Quiet mode
- KOR function
- KOR rate xx ml/h
- Date and time displaying
Selectable infusion mode:
- continuous rate
- easy pump
Syringe set configuration
Drug list configuration
Language configuration
Syringe size detector
Syringe plunger sensor
Power supply:
(220 - 230) VAC ±10%, 50/60 Hz, or internal rechargeable
battery
Power consumption:
10 VA (max)
Battery type:
Build-in NiMH 9.6 V 1300 mAh
Battery life:
8 h @ 5 ml/h
2 h @ 100 ml/h
Display type:
10 characters x 2 lines (LCD with back-light)
Protection:
Class II, CF, IPX4
CE0408 marked
FDA approved Council Directive 93/42/EEC concerning medical device
Electrical safety and EMC: EN 60601-(1, 1-2, 2-24)
e36
030006 marked: EN 1789:1999+A1:2003 for road ambulances
Options:
110 VAC ±10%, 50/60 Hz
12 VDC external supply
RS232
Nurse Call
Alarm volume adjustment
Battery NiMH 9.6 V 1800 mAh (battery life 16 h @ 5 ml/h)
Acceptance of 5 ml syringes
Weight:
2.6 kg
Dimensions:
135x305x195 cm
Warranty:
1 Year
|